:MEASure Commands
21
Keysight InfiniiVision M9241/42/43A PXIe Oscilloscopes SCPI Programmer's Guide
467
:MEASure:DEFine
Command Syntax
:MEASure:DEFine <meas_spec>[,<source>]
<meas_spec> ::= {DELay | THResholds}, for remaining syntax, see:
•
":MEASure:DEFine DELay Command Syntax"
•
":MEASure:DEFine THResholds Command Syntax"
The :MEASure:DEFine command sets up the definition for measurements by
specifying the delta time or threshold values. Changing these values may affect
the results of other measure commands. The table below identifies which
measurement results that can be affected by redefining the DELay specification or
the THResholds values. For example, changing the THResholds definition from the
default 10%, 50%, and 90% values may change the returned measurement result.
:MEASure:DEFine
DELay Command
Syntax
MEASure Command
DELay
THResholds
DUTYcycle
x
DELay
x
x
FALLtime
x
FREQuency
x
NWIDth
x
OVERshoot
x
PERiod
x
PHASe
x
PREShoot
x
PWIDth
x
RISetime
x
VAVerage
x
VRMS
x
NOTE
The DELay portion of the MEASure:DEFine command has been deprecated, and you should
instead use the :MEASure:DELay:DEFine command to specify delay measurement parameters.
The delay measurement now allows an edge occurrence of zero (0) to specify automatic edge
selection. The limitation of this command is that you cannot specify a negative slope with an
occurrence count of zero (0).
Содержание InfiniiVision M9241A
Страница 1: ...Keysight InfiniiVision M9241 42 43A PXIe Oscilloscopes SCPI Programmer s Guide ...
Страница 52: ...52 Keysight InfiniiVision M9241 42 43A PXIe Oscilloscopes SCPI Programmer s Guide 2 Setting Up ...
Страница 178: ...178 Keysight InfiniiVision M9241 42 43A PXIe Oscilloscopes SCPI Programmer s Guide 5 Commands Quick Reference ...
Страница 246: ...246 Keysight InfiniiVision M9241 42 43A PXIe Oscilloscopes SCPI Programmer s Guide 7 Root Commands ...
Страница 268: ...268 Keysight InfiniiVision M9241 42 43A PXIe Oscilloscopes SCPI Programmer s Guide 8 ACQuire Commands ...
Страница 280: ...280 Keysight InfiniiVision M9241 42 43A PXIe Oscilloscopes SCPI Programmer s Guide 9 CALibrate Commands ...
Страница 354: ...354 Keysight InfiniiVision M9241 42 43A PXIe Oscilloscopes SCPI Programmer s Guide 15 EXTernal Trigger Commands ...
Страница 412: ...412 Keysight InfiniiVision M9241 42 43A PXIe Oscilloscopes SCPI Programmer s Guide 17 FUNCtion m Commands ...
Страница 416: ...416 Keysight InfiniiVision M9241 42 43A PXIe Oscilloscopes SCPI Programmer s Guide 18 HCOPY Commands ...
Страница 442: ...442 Keysight InfiniiVision M9241 42 43A PXIe Oscilloscopes SCPI Programmer s Guide 20 MARKer Commands ...
Страница 588: ...588 Keysight InfiniiVision M9241 42 43A PXIe Oscilloscopes SCPI Programmer s Guide 23 MTESt Commands ...
Страница 692: ...692 Keysight InfiniiVision M9241 42 43A PXIe Oscilloscopes SCPI Programmer s Guide 24 POWer Commands ...
Страница 1062: ...1062 Keysight InfiniiVision M9241 42 43A PXIe Oscilloscopes SCPI Programmer s Guide 30 TIMebase Commands ...
Страница 1250: ...1250 Keysight InfiniiVision M9241 42 43A PXIe Oscilloscopes SCPI Programmer s Guide 33 WGEN w Commands ...
Страница 1306: ...1306 Keysight InfiniiVision M9241 42 43A PXIe Oscilloscopes SCPI Programmer s Guide 35 Obsolete and Discontinued Commands ...
Страница 1314: ...1314 Keysight InfiniiVision M9241 42 43A PXIe Oscilloscopes SCPI Programmer s Guide 36 Error Messages ...
Страница 1376: ...1376 Keysight InfiniiVision M9241 42 43A PXIe Oscilloscopes SCPI Programmer s Guide 39 More About Oscilloscope Commands ...