:FRANalysis Commands
16
Keysight InfiniiVision M9241/42/43A PXIe Oscilloscopes SCPI Programmer's Guide
369
:FRANalysis:WGEN:VOLTage
Command Syntax
:FRANalysis:WGEN:VOLTage <amplitude>,[<range>]
<amplitude> ::= amplitude in volts in NR3 format
<range> ::= {F20HZ | F100HZ | F1KHZ | F10KHZ | F100KHZ | F1MHZ
| F10MHZ | F20MHZ}
The :FRANalysis:WGEN:VOLTage command specifies the waveform generator's
output sine wave amplitude.
Use the :WGEN:VOLTage:OFFSet command to specify the offset voltage or DC
level.
You can also specify the amplitude and offset indirectly using the
:WGEN:VOLTage:HIGH and :WGEN:VOLTage:LOW commands. For example, an
amplitude of 5 V and an offset of 1 V is the same as a high-level voltage of 4 V and
a low-level voltage of -1 V.
When the amplitude profile setting is on (:FRANalysis:WGEN:VOLTage:PROFile
ON), the <range> option specifies the initial ramp amplitude at the frequency
setting. Amplitudes ramp between the settings specified for individual frequencies.
Query Syntax
:FRANalysis:WGEN:VOLTage? [<range>]
The :FRANalysis:WGEN:VOLTage? query returns the currently specified waveform
generator amplitude.
When the amplitude profile setting is on (:FRANalysis:WGEN:VOLTage:PROFile
ON), the <range> option specifies the frequency whose initial ramp amplitude is
returned.
Return Format
<amplitude><NL>
<amplitude> ::= amplitude in volts in NR3 format
See Also
•
":FRANalysis:WGEN:VOLTage:PROFile"
•
•
•
•
•
•
•
•
•
Содержание InfiniiVision M9241A
Страница 1: ...Keysight InfiniiVision M9241 42 43A PXIe Oscilloscopes SCPI Programmer s Guide ...
Страница 52: ...52 Keysight InfiniiVision M9241 42 43A PXIe Oscilloscopes SCPI Programmer s Guide 2 Setting Up ...
Страница 178: ...178 Keysight InfiniiVision M9241 42 43A PXIe Oscilloscopes SCPI Programmer s Guide 5 Commands Quick Reference ...
Страница 246: ...246 Keysight InfiniiVision M9241 42 43A PXIe Oscilloscopes SCPI Programmer s Guide 7 Root Commands ...
Страница 268: ...268 Keysight InfiniiVision M9241 42 43A PXIe Oscilloscopes SCPI Programmer s Guide 8 ACQuire Commands ...
Страница 280: ...280 Keysight InfiniiVision M9241 42 43A PXIe Oscilloscopes SCPI Programmer s Guide 9 CALibrate Commands ...
Страница 354: ...354 Keysight InfiniiVision M9241 42 43A PXIe Oscilloscopes SCPI Programmer s Guide 15 EXTernal Trigger Commands ...
Страница 412: ...412 Keysight InfiniiVision M9241 42 43A PXIe Oscilloscopes SCPI Programmer s Guide 17 FUNCtion m Commands ...
Страница 416: ...416 Keysight InfiniiVision M9241 42 43A PXIe Oscilloscopes SCPI Programmer s Guide 18 HCOPY Commands ...
Страница 442: ...442 Keysight InfiniiVision M9241 42 43A PXIe Oscilloscopes SCPI Programmer s Guide 20 MARKer Commands ...
Страница 588: ...588 Keysight InfiniiVision M9241 42 43A PXIe Oscilloscopes SCPI Programmer s Guide 23 MTESt Commands ...
Страница 692: ...692 Keysight InfiniiVision M9241 42 43A PXIe Oscilloscopes SCPI Programmer s Guide 24 POWer Commands ...
Страница 1062: ...1062 Keysight InfiniiVision M9241 42 43A PXIe Oscilloscopes SCPI Programmer s Guide 30 TIMebase Commands ...
Страница 1250: ...1250 Keysight InfiniiVision M9241 42 43A PXIe Oscilloscopes SCPI Programmer s Guide 33 WGEN w Commands ...
Страница 1306: ...1306 Keysight InfiniiVision M9241 42 43A PXIe Oscilloscopes SCPI Programmer s Guide 35 Obsolete and Discontinued Commands ...
Страница 1314: ...1314 Keysight InfiniiVision M9241 42 43A PXIe Oscilloscopes SCPI Programmer s Guide 36 Error Messages ...
Страница 1376: ...1376 Keysight InfiniiVision M9241 42 43A PXIe Oscilloscopes SCPI Programmer s Guide 39 More About Oscilloscope Commands ...