Section 4: Command reference
Series 2200 Programmable DC Power Supplies Reference Manual
4-24
2200S-901-01 Rev. C / March 2016
[SOURce:]VOLTage:PROTection[:LEVel]
This command sets the over voltage protection (OVP) threshold
level.
Group
Source
Syntax
[SOURce:]VOLTage:PROTection[:LEVel]
{<voltage>|MIN|MAX|DEF}
[SOURce:]VOLTage:PROTection[:LEVel]?
Related
Commands
[SOURce:]VOLTage:PROTection[:LEVel]
(on page 4-24)
(on page 4-26)
Arguments
<voltage>::=<NRf><units>
where:
<NRf>
is a flexible decimal that specifies the OVP threshold,
ranging from 0 to the maximum OVP level
<units> ::={V|mV}
MIN:
sets the minimum OVP level (1 V).
MAX:
sets the maximum OVP level, which is approximately 10%
higher than the maximum output voltage rating of the power supply.
DEF:
sets the OVP level to the default level, which is equal to MAX.
Returns
<NR2>
is the OVP threshold in volts.
MIN:
set to minimum OVP level.
MAX:
set to maximum OVP level.
Examples
VOLT:PROT 30V
VOLT:PROT?
might return 30, which would be the OVP threshold in
volts.