Series 2200 Programmable DC Power Supplies Reference Manual
Appendix B: Calibration
2200S-901-01 Rev. C / March 2016
B-29
8.
Change the AC Power Source output to the voltage specified in the following table.
DUT voltage selector switch
AC power source voltage
110
115 V
220
230 V
9.
Enter the voltmeter reading into the Nom line for your product in the table
(on page B-9).
10. Calculate the two values: Nom – Min and Max – Nom. Enter the values into the appropriate
(on page B-9) table.
11. Select the largest of the two calculations from step 12 and enter the value into the Largest column
in the
(on page B-9) table.
12. Turn the DUT output off.