Series 2200 Programmable DC Power Supplies Reference Manual
Appendix B: Calibration
2200S-901-01 Rev. C / March 2016
B-25
8. Enter the voltmeter reading (where OVP triggering occurred) into the
(on page B-6) table for checking DC Overvoltage Detection.
9. Reset OVP.
a. Press
Shift
and
Menu
(
1
).
b. Press the down arrow until
Protect
is shown and press
Enter
.
c. Press the down arrow until
Reset Protect
is shown and press
Enter
.
d. Press the right arrow to select
Yes
and press
Enter
.
e. Press the
Esc
key as needed to exit the menu system.