Series 2200 Programmable DC Power Supplies Reference Manual
Appendix B: Calibration
2200S-901-01 Rev. C / March 2016
B-3
DC voltage accuracy with remote sense
Instrument
DUT voltage
Test current
Min
Measured
Max
0% Test Voltage
2200-20-5
0 V
2.5 A
-0.0030 V
__________
0.0030 V
2200-30-5
0 V
2.5 A
-0.0030 V
__________
0.0030 V
2200-32-3
0 V
1.5 A
-0.0030 V
__________
0.0030 V
2200-60-2
0 V
1.25 A
-0.0060 V
__________
0.0060 V
2200-70-1
0 V
0.63 A
-0.0060 V
__________
0.0060 V
25% Test Voltage
2200-20-5
5 V
2.5 A
4.9955 V
__________
5.0045 V
2200-30-5
7.5 V
2.5 A
7.49475 V
__________
7.50525 V
2200-32-3
8 V
1.5 A
7.9946 V
__________
8.0054 V
2200-60-2
15 V
1.25 A
14.9895 V
__________
15.0105 V
2200-70-1
18 V
0.63 A
17.9886 V
__________
18.0114 V
50% Test Voltage
2200-20-5
10 V
2.5 A
9.9940 V
__________
10.0060 V
2200-30-5
15 V
2.5 A
14.9925 V
__________
15.0075 V
2200-32-3
16 V
1.5 A
15.9922 V
__________
16.0078 V
2200-60-2
30 V
1.25 A
29.9850 V
__________
30.0150 V
2200-70-1
36 V
0.63 A
35.9832 V
__________
36.0168 V
75% Test Voltage
2200-20-5
15 V
2.5 A
14.9925 V
__________
15.00750 V
2200-30-5
22.5 V
2.5 A
22.49025 V
__________
22.50975 V
2200-32-3
24 V
1.5 A
23.9898 V
__________
24.0102 V
2200-60-2
45 V
1.25 A
44.9805 V
__________
45.0195 V
2200-70-1
54 V
0.63 A
53.9778 V
__________
54.0222 V
100% Test Voltage
2200-20-5
20 V
2.5 A
19.9910 V
__________
20.0090 V
2200-30-5
30 V
2.5 A
29.9880 V
__________
30.0120 V
2200-32-3
32 V
1.5 A
31.9874 V
__________
32.0126 V
2200-60-2
60 V
1.25 A
59.9760 V
__________
60.0240 V
2200-70-1
72 V
0.63 A
71.9724 V
__________
72.0276 V