Section 4: Command reference
Series 2200 Programmable DC Power Supplies Reference Manual
4-22
2200S-901-01 Rev. C / March 2016
[SOURce:]OUTPut:TIMer[:STATe]
This command turns the output timer function on and off. When the
timer is activated and a duration is set, the output of the power
supply will turn off automatically if left on longer than the specified
duration. In order to ensure proper operation of the output timer, the
power supply must be turned on by using the
(on page 4-22) command.
Group
Source
Syntax
[SOURce:]OUTPut:TIMer[:STATe] {0|1|ON|OFF}
[SOURce:]OUTPut:TIMer[:STATe]?
Related
Commands
[SOURce:]OUTPut:PROTection:CLEar (no query form)
(on page 4-
(on page 4-21)
(on page 4-20)
Arguments
0
or
OFF
turns the output timer off.
1
or
ON
turns the output timer on.
Returns
0|1
Examples
To start timer, first send
OUTPUT:TIMER:STATE ON
, then send
OUTPUT:STATE ON
.
To end timer (turn timer off), send
OUTPUT:TIMER:STATE OFF
.