CAUTION
Risk of injury and material damage from incorrect use of the charger
for the wireless foot control.
Personal injuries, damage to the wireless foot control or the charger.
▶
Do not use the treatment unit during the charging process!
▶
Do not use the wireless foot control charger supplied to charge non-re-
chargeable batteries.
▶
Charge the wireless foot control with the charger supplied only.
Note
Charge the wireless foot control with the charger supplied by KaVo only.
Note
The wireless foot control charger may only be used indoors and must be pro-
tected from moisture.
▶
Connect the charger to the wireless foot control.
The charger display communicates the following:
Display
Meaning
shines green
The unit is ready
shines yellow
The battery is being charged
Shines weak green
Battery is charged
does not shine
The battery is dead, or there is a short-circuit
The battery voltage is above the tolerance range
Reversed poles
The transition from charging to full is indicated by the fluttering of the display.
4.10 Using instruments
Note
Consult the separate instructions for the installation, use and servicing of the
individual handpieces (e.g. turbine, COMFORTdrive, ultrasonic scaler, cam-
era, Satelec Mini LED, etc.) for pertinent information.
4.10.1 Holder logic
All handpieces on the dentist's side are secured against simultaneous use by
holder logic. If an handpiece has been removed when the unit is switched on,
the relevant holder will not be activated until the relevant handpiece has been
replaced.
Only the withdrawn handpiece is active, i.e., any handpiece that is withdrawn
afterward is not started. Exception: MF handpiece (parallel operation is possible
here).
4.10.2 Using suction hoses
▶
Remove the spray mist suction device or saliva ejector from the holder.
ð
The spray mist suction device or saliva ejector automatically turns on, and
then when it is placed in the holder, it turns off.