Instructions for use ESTETICA E50 Life
7 Safety check - Test instructions | 7.2 Instructions for the safety check
136 / 162
Note
The integrity of the power supply cable, in particular the protective earth wire
of the power cable must be ensured. As this is a fixed installation, the evalu-
ation can be conducted by means of a visual inspection. If damage is determ-
ined, the further procedure to be taken is specified in the general instruc-
tions.
Note
In this measurement the resistance of the protective earth connection of the
supply network can be taken into consideration.
Note
If applicable: all removable supply connection lines, which are kept handy for
possible use, should be taken into consideration and the respective PE meas-
ured.
Protective earth measurement
The protective conductor resistance must be measured at the following parts of
the device:
▪ Treatment centre
▪ Treatment lamp
▪ Accessories
Note
Additional measuring points SL X need to be taken into consideration in the
presence of accessories: e.g. additional devices, such as third-party con-
nector, USB port of the intraoral camera, etc.
See also:
2