![KaVo ESTETICA E50 Life Скачать руководство пользователя страница 118](http://html2.mh-extra.com/html/kavo/estetica-e50-life/estetica-e50-life_instructions-for-use-manual_3799123118.webp)
Instructions for use ESTETICA E50 Life
4 Operation | 4.12 Use pump for physiological saline solution (optional accessory)
118 / 162
+
-
g
u
a
S
k
c
u
r
D
N
a C
l
▶
Remove the old hose set sterile and dispose of it.
▶
Make sure that the flexing system is horizontal (turn manually if necessary).
Insert the new hose set sterile with tube holder from below, and turn the
pump tubing locking to the left.
+
-
g
u
a
S
k
c
u
r
D
N
a C
l
▶
Stick the puncture needle
⑥
into the coolant container and hook-in the
coolant container on the bottle holder.
▶
Check the sealing and firm seating of the puncture needle
⑥
. Prevent fluid
from leaking above the device.
▶
If you use a glass bottle, open the ventilation on the puncture needle
⑥
.
▶
If you use a bag, keep the ventilation on the puncture needle
⑥
closed.