Plug the second plug X2 of the KaVo measuring cable into the network card
(X2).
▶
Insert the protective contact plug of the KaVo measuring cable into the
safety tester.
Connecting the safety tester without the KaVo
measuring cable to the treatment centre
PE
N
L
▶
Switch L + N of the on-site power supply cord to be voltage-free.
▶
Disconnect L + N on terminals X1.1 and X1.2.
▶
Connect the safety tester directly to terminals X1.1 (L) and X1.2 (N) and
protective earth conductor terminal (PE).
Note
The main switch of the ME device / ME system must be turned on during the
measurement.
Connect the application parts [AP] to the safety tester: