Instructions for use ESTETICA E50 Life
9 Troubleshooting
9 Troubleshooting
Note
In case of malfunctions, consult the separate instructions for the use and
care of the individual instruments (such as the turbine, motor, camera, Sate-
lec Mini LED, etc.).
Malfunction
Cause
Remedy
Nothing works.
Main switch is off.
▶
Turn on main switch.
Main service fuse inter-
rupted the electric cir-
cuit.
▶
Unplug the unit from the mains.
▶
Check and replace, if required, the main ser-
vice fuse.
The main service fuse is situated next to the
master switch.
▶
For this purpose, open the bayonet closure
with a screwdriver and replace the fine-wire
fuse. (220, 230, 240 V AC: T 6,3 H
Mat. no.
0.223.2783
); (100, 110, 120, 130 V AC: T 10
H
Mat. no. 1.007.2529
).
▶
The re-close the bayonet closure with the
screwdriver.
The patient chair does
not move.
The safety shutoff is
activated.
(LED on the control
panel flashes.)
▶
Check the safety shutoff and eliminate the
reason for the shutoff.
Display without indic-
ator.
Bus / hardware error.
▶
Turn the device off and on.
▶
Call the service technician to look into the
problem if it continues to exist.
Operating device no
function.
Bus / hardware error.
▶
Turn the device off and on.
▶
Call the service technician to look into the
problem if it continues to exist.
Several handpieces are
simultaneously activ-
ated.
Hardware error.
▶
Stop working and call the service technician
The LED on the "LP/AP"
button flickers.
(Assistant unit)
The data connection to
the foot control is
faulty.
▶
Turn wireless foot control on.
▶
Call a service technician.
The LED on the "AP1"
button flickers.
(Dentist unit)
The data connection to
the assistant unit is
faulty.
▶
Turn wireless foot control on.
▶
Call a service technician.
The LED on the "AP2"
button flickers.
(Assistant element)
The data connection to
the chair control is
faulty.
▶
Call a technician.
Turbine making loud
running noises.
Turbine wheel faulty.
▶
Replace turbine wheels.
Follow the operating instructions for the tur-
bine.
151 / 162