8 Appendix - Additional measuring sites | 8.1 Additional scanning sites SL X in the protective conductor measure-
ment
148 / 162
8 Appendix - Additional measuring sites
Note
Regarding accessories, which are not listed here, comply with the specifica-
tions of the relevant instructions for use.
8.1 Additional scanning sites SL X in the protective
conductor measurement
ERGOcam One module
①
Screw in bottom part of housing
Note
The modules are not earthed with a safety conductor. In the case of excess-
ive PE resistance, the electrical connection between the module and the dent-
ist element must be improved. This can be accomplished, for example, by
means of a serrated lock washer on the fastening screw.
Connector for third-party equipment
▶
Position the test tip on the middle contact
①
.