4 Operation | 4.9 Operating the foot switch
▶
Press the "Blown air" foot button.
ð
As long as the footswitch is pressed, blown air exits the removed handpiece
(not with the PiezoLED)
4.9.10 Preselect counterclockwise motor rotation
▶
Take motor from the holder.
ð
The handpiece is active.
▶
Slide the cross switch upward.
ð
The direction of motor rotation is reversed each time the cross-switch is ac-
tuated: counterclockwise rotation - clockwise rotation.
ð
The direction of motor rotation is displayed on the dentist element.
4.9.11 Adjusting the instrument light
▶
Slide the cross switch to the right. (spotlight function)
ð
Cold light "On" (even when Cold light: "Off" is preselected).
▶
Slide the cross switch to the left.
ð
Change the cold light status: "On/Off"
4.9.12 Use physiological saline solution (optional
accessory)
Requirement
Treatment centre is turned on. The handpiece is connected to the pump via
the pressure line.
▶
Remove the handpiece.
▶
Press the cross-switch of the foot control for 4 seconds until you hear the
signal.
▶
After activation, select the "NaCl" cooling on the dentist control unit.