7 Safety check - Test instructions | 7.2 Instructions for the safety check
▶
Connect
①
to
④
to the safety tester.
▶
Connect the safety tester to additional measuring points AP X.
Note
Additional measuring points AP X must be taken into consideration in the
presence of accessories: e.g. accessories such as PIEZO ultrasonic scaler,
etc.
See also:
2
8 Annex - Additional measuring points, Page 148
Connect accessible conductive parts [ACP] with PE
ACP = accessible conductive parts
ACP
ACP
Note
Additional measuring points ACP X must be taken into consideration in the
presence of accessories.
See also:
2
8 Annex - Additional measuring points, Page 148
ACPs on the treatment centre
No ACPs need to be connected to the protective conductor (PE) during the
measurement on the treatment centre ESTETICA E50 Life, as all relevant parts
are connected to the PE at the factory and are included in the test.
ACPs on treatment lamps
No ACPs need to be connected to the operating lights during the measurement
with protective conductor (PE) because all relevant parts have already been
connected to the protective conductor (PE) at the factory and are included in
the test.
Measure protective conductor resistance
Limit
< 0,3 Ω (maximum value!)