Instructions for use ESTETICA E50 Life
8 Appendix - Additional measuring sites | 8.3 Additional connection sites ACP X (additional earth connections)
150 / 162
Note
The switch on the handpiece must be activated during the EPA measure-
ment.
Note
Additional measuring points AP X need to be taken into consideration in the
presence of accessories: e.g. if third-party devices are connected, camera of
the multimedia system, etc.
8.3 Additional connection sites ACP X (additional earth
connections)
Scan the saline pump with the test tip
+
-
g
u
a
S
k
c
u
r
D
N
a C
l
①
Test tip on bottle holder
Note
A fixed connection from ACP to the protective earth conductor terminal (PE)
must be established for the EUL and EPA measurement. This can be accom-
plished, e.g., using a measuring cable and connection terminals.