Instructions for use ESTETICA E50 Life
2 Safety | 2.3 Safety instructions
18 / 162
CAUTION
Health damage due to germ formation.
Infection hazard.
▶
Before starting, rinse all the water drain lines with no handpieces attached.
▶
Before start-up and after downtimes (weekends, holidays, vacations, etc.),
rinse or purge the air and water lines with air.
▶
Option: perform intensive germ reduction (if assembly kit is available).
▶
Actuate the tumbler filler several times.
CAUTION
Third-party device connection assembly kit (optional): Risk of reverse
contamination by stagnant water.
Infections.
If a water consumer is connected to the third-party device connection as-
sembly kit, always perform the following actions on the device:
▶
Before starting, rinse all the water drain lines with no handpieces attached
(if applicable).
▶
Before start-up and after downtimes (weekends, holidays, vacations, etc.),
rinse or purge the air and water lines with air.
▶
Mind the H
2
O
2
resistance of the water consumer since the water is doped
with OXYGENAL 6 (concentration up to 0.02 %).
CAUTION
Long stay in the patient chair.
Decubitus formation.
▶
Take precautions against the formation of decubitus in long treatments.
CAUTION
Risk of injury and material damage from incorrect use of the charger
for the wireless foot control.
Personal injuries, damage to the wireless foot control or the charger.
▶
Do not use the treatment unit during the charging process!
▶
Do not use the wireless foot control charger supplied to charge non-re-
chargeable batteries.
▶
Charge the wireless foot control with the charger supplied only.
CAUTION
Risk of injury when the dental chair or headrest is moved.
Hair of the patient or practice personnel may get caught when the headrest of
the dental chair is moved.
▶
Mind the hair of the patient or practice personnel when moving the dental
chair or the headrest.
CAUTION
Risk of injury when moving the patient or patient chair.
The patient or treatment personnel can be pinched or crushed.
▶
Position all moving parts, such as dentist element, assistant element, oper-
ating light, screens, etc., outside the collision range when you move the pa-
tient or patient chair.
NOTICE
Damage to the instrument hoses from stickers.
Instrument hoses can burst.
▶
Do not affix stickers or adhesive tape.