![KaVo ESTETICA E50 Life Скачать руководство пользователя страница 130](http://html2.mh-extra.com/html/kavo/estetica-e50-life/estetica-e50-life_instructions-for-use-manual_3799123130.webp)
Instructions for use ESTETICA E50 Life
7 Safety check - Test instructions | 7.2 Instructions for the safety check
130 / 162
Without/with DEKAmat/CENTRAmat
7.2.2 Visual inspection (inspection by examination)
Check the following items in advance:
▪ Has the equipment of the ME device or the ME system been changed since
the last inspection?
▪ Has the change been documented and approved (test protocol of safety
check STK)?
▪ Are there any indications of insufficient safety?
Check the ratings of fuses that are accessible from
outside
Note
A check of whether or not the main fuse on the mains socket corresponds to
the rated values is no longer included in DIN EN 62353:2015 [IEC
62353:2015].