7 Safety check - Test instructions | 7.2 Instructions for the safety check
142 / 162
Applied part leakage current - Alternative measuring
method
Limit
< 5 mA (maximum value!)
ACP
Protection class 1
WARNING
Electrical power.
Death or injury from electric shock.
▶
Conduct test for leakage current in devices of Protection Class I only after
the protective earth test has been passed.
WARNING
Electrical power.
Death or injury from electric shock.
▶
Prior to connecting the treatment centre to the safety tester, disconnect the
treatment centre from the mains supply network.
Note
In the testing of ME devices with several application parts, the parts must be
connected in succession. The measured results must be evaluated using the
threshold values. Application parts, which are not included in the measure-
ment, remain open.
Note
An additional measurement of the leakage current from type B applied parts
need only be carried out if this is specified by the manufacturer (see accom-
panying documents).
Note
A separate measurement is not usually required for type B applied parts. The
applied parts are connected to the casing (see diagram) and included in the
measurement of the leakage current of the casing, whereby the same per-
missible values are applicable.
7.2.4 Functional test
The following conditions must be met in all function tests:
▪ The basic function of the treatment centre must be guaranteed.
▪ The treatment centre must be fit for use.
▪ There must be no irregularities, noise or abrasion, etc., present.