Touch monitor with test tip:
▶
Scan the measuring point
①
with the test tip.
or
▶
Scan the measuring point
②
after removing the display cover.
Measure protective conductor resistance of accessories
See also:
2
8 Annex - Additional measuring points, Page 148
Equipment leakage current - Alternative measuring
method
Limit
< 10 mA (maximum value!)
ACP
Protection class 1
WARNING
Electrical power.
Death or injury from electric shock.
▶
Conduct test for leakage current in devices of Protection Class 1 only after
the protective earth test has been passed.
WARNING
Electrical power.
Death or injury from electric shock.
▶
Prior to connecting the treatment centre to the safety tester, disconnect the
treatment centre from the mains supply network.