Attachment locations: nameplate and note, "Comply with the instructions for use"
Control of the availability of the necessary documents
▶
Check if the required instructions for use and care instructions are available
in the surgery.
Note
Any irregularities determined in the visual inspection must be recorded in the
test protocol. It is essential to determine whether defects and deficiencies
could have an adverse impact on the safe operation of the unit. If the de-
termined irregularities present a safety hazard and cannot be rectified dir-
ectly, the unit must be closed down until safe operation is restored.
7.2.3 Measurements
WARNING
Danger to persons due to a lack of care exercised during the safety
checks and testing.
▶
Prior to connecting the treatment centre to the safety tester, disconnect it
from the mains supply network.
▶
Carry out all safety checks and tests in a manner that will ensure that there
will be no danger to the testing personnel, patients or other persons.
Note
The safety tester must comply with the requirements defined in DIN EN
62353 [IEC 62353], Annex C.
Note
If no other specifications have been made, all values relating to voltage and
current are effective values of alternating voltage, direct voltage or pulsating
voltage res. alternating current, direct current or pulsating current.