Kamstrup
A/S
∙
Technical
description
∙
5512
‐
2029_A1_GB_06.2017
7.9
Info
logger
Every
time
the
information
code
changes,
the
below
‐
mentioned
registers
are
logged.
Register
type
Description
Date
(20YY.MM.DD)
Year,
month
and
day
of
logging
time.
Clock
(hh.mm.ss)
Time
Info
Information
code
on
above
date
E1
Heat
energy
E3
Cooling
energy
It
is
possible
to
data
read
the
latest
250
changes
of
the
information
code
as
well
as
the
dates
the
changes
were
made.
When
the
information
code
is
read
from
the
display,
50
changes
including
dates
can
be
read.
All
of
the
250
changes
can
also
be
read
by
means
of
the
PC
program
LogView
HCW.
INFOEVENT
Every
change
of
a
parameter
of
the
info
code
results
in
an
info
event
if
the
selected
country
code
is
configured
with
the
parameter.
It
is
therefore
not
certain
that
all
parameters
result
in
an
info
event.
An
info
event
results
in
accumulation
of
the
info
event
counter
as
well
as
logging
in
the
info
logger.
This
does
not
apply
as
long
as
the
meter
is
in
transport
state
or
if
the
calculator
top
and
base
are
physically
separated.
Reconfiguration
of
active
parameters
of
the
info
code
will
influence
future
info
codes,
whereas
all
logged
info
codes
remain
as
they
were
at
the
time
of
logging.
7.10
Config
data
logger
Every
time
the
configuration
is
changed,
the
below
‐
mentioned
register
types
are
logged.
It
is
possible
to
data
read
the
latest
50
changes
of
the
config
log
as
well
as
the
dates
the
changes
were
made.
The
meter
permits
only
50
changes,
unless
you
break
the
legal
seal
and
carry
out
a
total
reset
of
the
meter,
which
also
resets
the
config
log.
Note:
The
fiftieth
change
of
configuration
must
be
carried
out
on
the
installation
site,
i.e.
either
via
the
SETUP
loop
or
via
METERTOOL
HCW.
Register
type
Description
Date
(20YY.MM.DD)
Year,
month
and
day
of
change
Hour
(hh.mm.ss)
Time
Configuration
number
The
new
configuration
number
E1,
E3
and
V1
Meter
counts
are
saved
just
after
reconfiguration
Hour
counter
Hour
counter
is
saved
t
offset
The
temperature
offset
value
is
saved.
V1
pulse
figure
The
pulse
figure
of
V1
(imp/l
or
l/imp)
is
saved
V1
q
p
Nominal
flow
q
p
is
saved
The
meter
will
always
carry
out
a
config
logging
if
the
user
has
had
access
to
the
SETUP
loop,
no
matter
whether
the
user
has
changed
the
configuration
or
not.