23.
Pre
‐
setting
t5
The
temperature
value
t5
can
be
set
in
the
interval
0.01…185.00
°C
in
the
SETUP
loop.
This
value
is
used
in
connection
with
the
calculation
of
outlet
energy
registers,
i.e.
registers
A1
(A
‐
,
heat
at
a
discount)
and
A2
(A+,
heat
with
an
addition).
Read
more
about
this
calculation
and
function
in
paragraph
7.1.3.
24.
EndSetup
The
menu
item
“EndSetup”
enables
the
technician
to
lock
the
access
to
the
SETUP
loop
in
transport
state
and
thus
lock
the
meter
against
further
configuration.
To
do
so,
the
user
must
keep
the
primary
key
activated
for
5
seconds.
During
the
five
seconds
the
frames
around
the
reading
EndSetup
will
currently
become
illuminated
in
the
meter’s
display.
This
action
can
be
undone
by
releasing
the
primary
key
before
the
whole
frame
has
become
illuminated,
i.e.
before
the
5
seconds
have
passed.
“EndSetup”
is
an
important
function
when
the
meter
is
in
transport
state,
but
when
the
meter
is
in
operation,
“EndSetup”
is
just
one
of
three
ways
to
exit
the
SETUP
loop.
See
paragraph
6.3.