MULTICAL®
603
102
Kamstrup
A/S
∙
Technical
description
∙
5512
‐
2029_A1_GB_06.2017
Below
are
examples
of
display
readings
of
the
date
and
data
readings.
The
frames
around
the
date
indicate
whether
the
date
is
connected
to
yearly
or
monthly
data.
In
data
readings
of
yearly
data
the
last
two
digits
of
the
year
are
marked
in
frames,
whereas
the
two
digits
for
month
are
marked
for
monthly
data.
Furthermore,
the
symbols
”MIN”
and
”MAX”
are
illuminated
to
indicate
whether
the
values
are
minimum
or
maximum
values.
The
symbol
“LOG”
is
illuminated
for
historical
readings.
Examples
of
yearly
date
and
data
(max
values)
for
flow
Date
of
max
this
year
Data
of
max
this
year
Date
of
max
yearly
logger
Data
of
max
yearly
logger
Examples
of
monthly
date
and
data
(min.
values)
for
power
Date
of
min.
this
month
Data
of
min.
this
month
Date
of
min.
monthly
logger
Data
of
min.
monthly
logger
All
minimum
and
maximum
values
are
calculated
as
the
average
of
a
number
of
current
flow
or
power
measurements
depending
on
the
chosen
averaging
period.
All
calculated
flow
and
power
values
from
the
averaging
period
are
used
in
the
calculation
of
the
average.
Calculated
values
are
compared
to
previous
values,
and
the
new
value
is
saved
if
it
exceeds
the
previous
maximum
value
or
is
lower
than
the
previous
minimum
value.
The
averaging
period
used
in
all
calculations
can
be
selected
in
the
interval
1...1440
minutes
in
leaps
of
1
minute
(1440
min.
=
24
hours).
The
averaging
period
and
the
target
dates
are
stated
in
the
order.
Read
more
about
order
data
in
paragraph
3.3.
If
not
otherwise
stated
in
the
order,
the
default
averaging
period
of
60
minutes
is
used.
This
value
can
later
be
changed
via
the
SETUP
loop
or
METERTOOL
HCW.
Please
note
the
following:
In
the
display
the
date
is
shown
in
the
format
20YY.MM.DD,
but
by
serial
reading
the
time
can
be
stated
too,
and
the
format
then
becomes
YY.MM.DD,
hh.mm.ss.
The
average
is
calculated
continuously
over
time,
i.e.
the
average
of
values
is
calculated
from
now
on
and
back
in
time
according
to
the
chosen
averaging
period.
As
a
result,
the
min/max
calculation
is
immune
to
the
clock
setting
and
will
always
move
continuously
through
time.