5.8 Ultrasonic probe check before and after operation
Check if there are any exceptionally on the surface of the probe or cable jacket, such as peeling, cracks,
bulge, or if the acoustic lens is reliable, disinfected or cleaned.
5.9 Main unit check before and after operation
5.9.1 Inspection before start-up
Check the following items before starting the machine:
1. The temperature, humidity and atmospheric pressure shall meet the requirements of operation
condition.
2. No condensation occurs.
3. No distortion, damage or contamination on system and peripheral. Clean the parts as specified in
relevant sections, if the contaminant is present.
4. Check the control panel, LED screen and enclosure to ensure they are in good working condition and
free of abnormity (such as cracks and loosened screws).
5. No damage on cable (power cable, etc.), and hard up on its connection.
6. Check probe and its connections to ensure they are free of abnormity (such as scuffing, drop-off or
contamination). If the contaminant is present, clean, disinfect the contaminated objects as specified
in relevant sections.
7. See to it that there are no foreign objects on the control panel (Remove the object if any).
8. No barriers around the intake of equipment.
9. See to it that probe has been cleaned, disinfected; else dispose it as specified in relevant sections.
10. Check all the ports of the machine for possible damage or blockage.
11. Clean the field and environment.
5.9.2 Inspection after start-up
Check the following items after starting the machine:
1. No abnormal voice, strange smell and overheating appear.
2. Check the machine to ensure a normal start-up: The power indication light is on and startup picture is
shown on the screen. Then the machine will automatically enter B mode.
3. Check the acoustic lens for abnormal heat when the probe is in use. This can be done by hand touching
the probe to feel the temperature of the lens.
4. Check the image to ensure trouble-free display (no excessive noise or flicker).
5. Check the control panel to ensure normal operation condition.
6. Check the instrument to ensure that the phenomenon of local high temperature will not appear.
Attention:
If the overheat acoustic lens is placed on the patient’s skin, heat injury may
occur.
Attention
: Thoroughly clean the coupling gel on the probe surface each time after
ultrasonic operation, or the coupling gel may become hardened on the acoustic lens of the
probe, deteriorating quality of image.