Full Digital Color Doppler Ultrasonic Diagnostic Instruments (Vet) User’s Manual V1.01
- 73 -
Chapter 13 Puncture guide
13. 1 Operation of puncture guide line
1. In the real time B mode, press the
Biopsy
key on the control panel, the Biopsy menu is displayed at the
top left of the screen.
2. Press
Cursor
key, move the cursor to [BGL Type] on the Biopsy menu, press the
Set
key to select the
BGL-1, BGL-2 or BGL-3.
Move the cursor to [Visiable], press the
Set
key to turn on or turn off the puncture guide line.
Move the cursor to [Position], press the
Set
key to set the position of the puncture guide line BGL-1,
BGL-2 or BGL-3.
Move the cursor to [Angle], press the
Set
key to set the angle of the puncture guide line BGL-1,
BGL-2 or BGL-3.
Move the cursor to [Save parameter], press the
Set
key to save the parameters.
Move the cursor to [Restore factory], press the
Set
key to restore to the factory.
3. Press the
Biopsy
key to exit the biopsy states.
Attention:
During the operation of puncture guide, do not freeze the image.