Full Digital Color Doppler Ultrasonic Diagnostic Instruments (Vet) User’s Manual V1.01
- 79 -
6
Dark image
1.
Check probe being used is connected to main unit well
2.
Adjust brightness
3.
Check whether the TGC sliders are at the left side, if yes,
please drag them to the right side
7
Control panel malfunction
Shut down the machine and restart it again.
14.7 Maintenance by service personnel
To ensure the performance and safety of instrument, it must be checked after using 1 year. When check
the instrument, please consult the International Trade Dept of Kaixin or its dealers, as they need to have
professional technology engineers.
The following safety checks should be performed at least every 12 months by a qualified person who
has adequate training, knowledge, and practical experience to perform these tests.
Inspect sorts
Inspect items
Cleaning
System inside,
Peripheral equipment
Electrical safety
Earth resistance,
Earth leakage current,
Enclosure leakage current
Patient leakage current,
Dielectric strength
Mechanical safety
Inspect the control panel,
Installed machine of peripheral equipment
Other mechanical part,
Probe appearance
Image recording
Each mode images,
Recorded images using typical probe.