5.1 System installation
Please carefully read through and fully understand the safety cautions before install and place the
system.
1. Unpack the instrument case and check the goods for its completeness according to the packing list.
2. Place the instrument on a stable and leveled position.
3. Put the ultrasonic couplant holder on the main unit and fasten it with the buckle; then screw the probe
cable hanger to the corresponding hole and tighten it.
4. Leave adequate space of 20 centimeters as minimum from rear, left and right side of the instrument.
5.2 Ultrasonic probe installation
Attention:
Adequate space from rear, left and right side of the machine shall be reserved,
or the machine may malfunction under excessive heat inside the enclosure.
Warning:
1. Do not use the probe not provided by our company, otherwise the equipment and the
probe will cause damage, and may cause fire in extreme cases.
2. Check the ultrasonic probe and connecting cable after diagnostic operation. Use of
defective probe may cause electric shock.
3. Do not strike the probe; using the damaged probe may cause electric shock to the patient.
4. Unauthorized disassembly of the probe shall be prohibited as it may cause electric shock.
Warning:
The probes and foot switch as described in this section may be replaced by
operator. To replace according to the method of this section, does not result in an
unacceptable risk.
Warning:
Only the peripherals and accessories (such as probes, peripherals or cables)
provided or recommended by the manufacturer can be used. Using other devices or
accessories may degrade the system performance and even cause an electrical shock.
Warning:
1.
Please make sure that the system is powered off and the power cable and the relevant
accessories are disconnected before moving or transporting the system.
2.
Do not place the system on a tilted plane with the angle larger than 10
°
. Otherwise, the
systemwill fall off to cause system damage or personal injury.
3.
The plug is used as disconnect to the mains supply, do not to position the machine so