Full Digital Color Doppler Ultrasonic Diagnostic Instruments (Vet) User’s Manual V1.01
- 14 -
5.2.1 Connecting the probe
Steps for connection are as follows.
1. Shutdown the system, place the probe cable downward and rotate the probe lock on the probe
connector to the horizontal direction, as shown below.
2. Insert the connector into the probe port until squeeze it.
3. Rotate the probe lock on the probe connector to vertical direction position to make it fixed, as shown
below.
5.2.2 Disconnecting the probe
Shutdown the system, rotate the probe lock on the probe connector to the horizontal direction and
remove the probe connector, as shown below.
5.3 Connect to equipotential wire
Shutdown the system, connect one end of the equipotential wire to the equipotential terminal (
)
and
the other end to the earth.
Attention:
1. Turn off the ultrasonic system before disconnecting the probe. Disconnecting the probe
with system power on may damage the system or probe.
2. Before disconnecting the probe, place the probe on the probe holder and put its cable
on the probe cable hanger so that the probe may not be damaged or injury person by
unexpected fall.
3. Freeze the instrument when instrument is start-up without operation to increase of
service life of probe.
4. Repeat available machine time should be more than one minute to avoid turn on/off
power supply in short time.
5. Disconnect the probe from the system after freezing an image or powering off the
system. Otherwise, the system or the probe could be damaged.
Warning:
Before connecting or using the probe, make sure that the probe, connecting
cable and connector are in normal condition (free of cracks or drop). Use of defective
probe may cause electric shock.
Warning:
Equipotential wire used in the system shall conform to the requirements in
the IEC 60601-1.
Warning:
Disconnect the probe when the image is frozen or the system is shut down.
Otherwise, may damage the probe and system.