Full Digital Color Doppler Ultrasonic Diagnostic Instruments (Vet) User’s Manual V1.01
- 11 -
Chapter 4 Operation Condition
4.1 Power supply
Power supply: 100-240V~ 50/60Hz
,
1.2-0.6A
4.2 Operation Environment
Ambient temperature: 10
℃
-40
℃
Relative humidity: 30%-75% (without condensation)
Atmospheric pressure: 700hPa-1060hPa
Altitude: < 2000 m
Overvoltage: Overvoltage Category
Ⅱ
Pollution degree: 2
4.3 Storage and Transport
Ambient temperature: -20
℃
-55
℃
Relative humidity: 30%-93% (without condensation)
Atmospheric pressure: 700hPa-1060hPa
Attention:
The main voltage is varies with different countries or regions.
Danger:
Do not use this equipment where flammable gas (such as anesthetic gas, oxygen
or hydrogen) or flammable liquid (such as alcohol) are present. Failure to do so may
result in explosion.
Warning:
Avoid using this equipment with high-frequency electric knife,
high-frequency therapy equipment or defibrillators and other electronic devices, or may
an electric shock occur to the patient.
Attention:
Using radio transmitting equipment nearby the system may interfere with
the normal operation of the system. Prohibited carry or use of devices that can generate
radio waves within the room installed this system, such as cell phones, radio transceivers
and wireless remote control toys.
Attention:
System should be avoided using in following environments:
1. Splash
2. Moist
3. Rain
4. Thunderstorm weather 5. No ventilation
6. Dust
7. Close to heat source
8. Direct sunlight 9. Dramatic temperature change
10. Chemical medicines
11.Poisonous gas 12. Corrosive gas
13. Strong shock
14. Strong electromagnetic field (e.g. MRI)
15. Radiation (e.g. X-ray, CT) 16. Defibrillators or short wave therapy equipment