Full Digital Color Doppler Ultrasonic Diagnostic Instruments (Vet) User’s Manual V1.01
- 16 -
2. Connect the printer to the mains supply with power cable, with AV video cable connect composite
video output interface (
) of the system to the printer video input interface. You can use the video
printer when the connection is completed.
5.6.3 Connecting the graphic printer
Please use the printer model recommended by the manufacturer to connect the instrument. The
graphic printer is purchased by the user.
Shutdown the system, connect the printer to the mains supply with power cable, then connect the
instrument and printer by USB cable. You can use the graphic printer when the connection is completed and
set the graphic printer model in the System Preset.
5.7 Booting/Standby/Powering off
Booting the system
Press the
Power on/Standby
key in the upper-left corner of the control panel to boot the system.
Note: The probe must be connected before power on.
Standby
In standby mode, screen is closed, system is in low power status, the
Power on/Standby
key is
blinking in green backlight.
Go to sleep:
Press the
Power on/Standby
key in the upper-left corner of the control panel, and select the
[Standby] to enter the standby mode.
Press
Preset
key
[System Preset]
[Standby] and [Wait]. The system enters the standby mode
automatically when the user does not perform any operaton within the set time.
Go to sleep:
Press the
Power on/Standby
key in the upper-left corner of the control panel to exit the standby
mode.
Powering off the system
Press the
Power on/Standby
key in the upper-left corner of the control panel, and select the [Shut off]
to shut down the system.
Attention:
1.
By using factory default settings, according to actual situation, please adjust the default
parameters of graphic printer to get the best quality printed image.
2.
Please check user instruction to install the printer. If the printer doesn't work, please
press Preset key and choose [System Preset] to check related printer preset.
3.
The system supports composite video output, video mode can select NTSC or PAL.
Attention:
1. The instrument can not be powered on again after the power is turned off for one
minute or it may be faulty.
2. Do not power off the system during system upgrade or data transmission.
3. Do not disconnect the power cable before the power off information disappears.
Otherwise, the files may be damaged and patient data may be lost.
4. For safety and functionality of the system, regular maintenance should be performed
as described in the section "System Maintenance".
5. Long-press the Power on/Standby key, you can turn off the system forcibly but may
cause data damaged or lose, avoid turn off the system in this way unless it is necessary.
6. When the device crashes, press the Power on/Standby key in the upper-left corner of
the control panel to restart the system.