HiSeq X System User Guide
69
Stagger Runs on Flow Cell A and Flow Cell B
1
Wait for the run on the adjacent flow cell to begin a chemistry step, and then select
Pause
. The Pause menu opens.
NOTE
Always pause the current run during a chemistry step opposed to an imaging step.
2
Select
Normal Pause
.
3
Wait for the software to complete the current chemistry step. The system is placed in a
safe state automatically.
4
After the adjacent flow cell run is paused, set up the new run.
5
After loading the new flow cell for the new run, close the compartment door.
6
Select
Start
to start the new sequencing run.
7
Select
Resume
on the adjacent flow cell to resume the paused run. The software
automatically controls chemistry and imaging processes on both flow cells.