2-11
2
2.7
CONFIGURING WITH AN INTERACTIVE PROGRAM
2.7.1 General Configuration Instructions
The 4899A can be easily configured from any GPIB bus controller with
an interactive program such as ICS's GPIBkybd program by using the fol-
lowing procedure.
1.
Connect the unit to the GPIB controller as shown in Figure
2-2. Run the program and if required use an Abort, REN or a take
control type command to have the bus controller assert REN. Then
turn the unit on.
2.
Determine the unit’s GPIB address:
a)
For new units use the factory setting of 04.
b)
For other units, turn the unit off and back on. At the
end of its self test, the unit blinks its GPIB address on the LEDs.
RDY TALK LSTN SRQ ERR
16
8
4
2
1
Figure 2-4 GPIBkybd Program with 4829A's IDN response