5 Troubleshooting
40
Tip
ON : Activates as configured.
Output error ON: Activates as output error. (Connected between C-NC)
OFF: Alarm contact output behavior stops.
Hold: Outputs the last value.
5.4
Message code
This device has a function to display various message codes. Message codes are displayed
at the bottom of the display.
5.4.1
Description and action for message code
When message are displayed, perform the operation in accordance with the following table.
Note
If the message code is still displayed after performing the action in the table, contact us.
High Pressure
Err
Output error
ON
Hold
Red flash
YES
YES
Hold
Output error
ON
TOC
OFF
Cond
ON
Temp
ON
Error code
Control
target
Alarm contact
output
Transmittance
output
Display lamp
Error code display
Maintenance
OFF
Maintenance
ON
Message code
Description
Cause or possible cause
Action
Warming Up
During warm-up
Function configured at “con-
figuration of warming up
time” is functioning.
Wait for a while until warming up
function completes.
UV Total Over
UV lamp integration time
error
UV lamp total time exceeds
8800 hours.
Replace the UV lamp and clear
the total time.