
2 System Overview
11
Q
Setting menu
The stage under the maintenance mode includes the following menu.
For detailed step to enter each menu and parameter, refer to the description in Table 1.
z
System setting (System Set)
z
Measurement condition setting (Meas. Set)
z
Output setting (Output Set)
Table 1 Basic setting parameters
Display
Description
Selection
Default
Reference
System Set Menu
System setting
Page 24
Mean.Stop Set
Setting for when measurement is stopped using external input
Page 27
UV Off
Setting of UV lamp operation when mea-
surement is stopped
Yes, No
Yes
Page 27
Delay
Setting of the amount of time that will
elapse between input and the point that
the UV lamp is turned OFF when the
"Setting of UV lamp operation" has been
set to "Yes"
1 to 120 (min)
5
Page 27
Warm Up
Setting of the amount of warm-up time
when external input is reset
1 to 60 (min)
30
Page 27
No Sample Set
Setting for no sample detection time
Page 25
UV Off
Setting of UV lamp operation during no
sample detection time
Yes, No
Yes
Page 25
Delay
Setting of the amount of time that will
elapse between the point of no sample
detection and the UV lamp is turned OFF
when the "Setting of UV lamp operation"
has been set to "Yes"
1 to 120 (min)
5
Page 26
Warm Up
Setting of the warm-up time when the no
sample detection has been resolved
1 to 60 (min)
30
Page 26
UV Lamp Set
Setting for the UV lamp
Page 27
Total
Confirmation and resetting of the total UV lamp time
Page 27
Meas. Set Menu
Measurement condition setting
Page 28
Range
Measurement range selection
1000, 500.0, 100.0 (ppb)
1000
Page 28
M.AVG Set
Moving average count setting
Page 28
TOC
TOC moving average count setting
1 to 20 (count)
10
Page 28
Temp
Temperature moving average count
1 to 20 (count)
10
Page 29
Cond
Conductivity moving average count
1 to 20 (count)
10
Page 29
Err. Option Set
Optional setting for error
Page 30
Cond Over
Input conductivity error selection
Yes, No
Yes
Page 30
Temp Over
Sample temperature error selection
Yes, No
Yes
Page 31
Output Set Menu
Output setting
Page 32
Alarm Set
Alarm contact output setting
Page 32
Alarm1
Alarm contact output setting1
Err, Hold, TOC, Cond, Temp Err
Page 32
Alarm2
Alarm contact output setting2
TOC (Fixed)
TOC
Page 34
Analog Set
Transmittance output setting
Page 35
Span
Span point setting
Depends on the set “Range”
of measurement range.
1000
Page 36