4 Measurement
34
Alarm contact output setting 2 (Alarm2)
1.
Perform the procedures 1. and 2. in "
Q
Option setting for errors (Err. Option Set)"
(page 30).
2.
Select “Alarm2” and press the ENT key.
The screen display “Alarm2 Set Menu” and the alarm contact output setting 2 can be set.
Tip
The control target for "Alarm contact output setting 2 (Alarm2)" is fixed at "TOC".
Control value (Set)
Set the control value as a standard for the point where the alarm contact output is turned ON
and OFF.
1.
Perform the procedures 1. and 2. in "
Alarm contact output setting 2 (Alarm2)"
(page 34).
2.
Select “Set” and press the ENT key.
The display flashes to show now the setting can be performed.
3.
Press the
▲
/
▼
key to set the control value.
▲
key: Value increases by the minimum display unit.
▼
key: Value decreases by the minimum display unit.
Setting range is 0 to the set value in “Selecting measurement range”.
4.
Press the ENT key to confirm the setting.
Delay time (when the alarm contact output is ON) (Delay)
Set the delay time from when the alarm is output until the alarm contact output setting 2 is
operated.
1.
Perform the procedures 1. and 2. in "
Alarm contact output setting 2 (Alarm2)"
(page 34).
2.
Confirm “Delay” is selected and press the ENT key.
The display flashes to show now the setting can be performed.
3.
Press the
▲
/
▼
key to set the control value.
▲
key: Value increases by 1 second.
▼
key: Value decreases by 1 second.
Setting range is 0 seconds to 60 seconds.