![horiba HT-110 Скачать руководство пользователя страница 26](http://html1.mh-extra.com/html/horiba/ht-110/ht-110_instruction-manual_2162223026.webp)
3 Installation
18
In case of 4 mA to 20 mA DC
When connecting several receiving instruments
Connect in series as follows.
Resistance of receiving instruments should be less than 600
Ω
in total.
In case of 0 V to 1 V DC (optional)
When connecting several receiving instruments
Connect in parallel as follows.
Parallel combined resistance of the termination resistance 1 and 2 should be over 100 K
Ω.
Reception resistance
less than 600
Ω
OUTPUT (mA)
(+)
(
−
)
FG
Receiving instrument
Shield
OUTPUT (mA)
(+)
(
−
)
Receiving instrument 1
Receiving instrument 2
Reception
resistance 1
Reception
resistance 2
Input impedance over
100 k
Ω
Receiving instrument
OUTPUT (V)
(+)
(
−
)
FG
Shield
OUTPUT (V)
(+)
(
−
)
Input impedance 1
Receiving instrument 1
Receiving instrument 2
Input impedance 2