4 Measurement
26
Delay time setting (Delay)
Set the amount of time that will elapse between the point of no sample detection and the UV
lamp is turned OFF when the "Setting of UV lamp operation" has been set to "Yes".
1.
Perform the procedures 1. and 2. in "
Q
Setting for no sample detection time
setting(No Sample)" (page 25).
2.
Select "Delay" and press the ENT key.
Setting is now possible and the display flashes.
3.
Use the
▲
and
▼
keys to select the desired operation for the UV lamp.
▲
key: Value increases by 1 minute.
▼
key: Value decreases by 1 minute.
The setting range is from 1 minute to 120 minutes.
4.
Press the ENT key to register the setting.
Tip
z
This setting is only available when the "Setting of UV lamp operation" has been set to "Yes".
(We recommend the delay time setting to 5 minutes)
z
When pressing the MENU key, the menu goes back to the previous menu.
Warm-up time setting (Warm Up)
Set the amount of time that will elapse from the point that measurement is stopped to the
beginning of measurement.
1.
Perform the procedures 1. and 2. in "
Q
Setting for when measurement stops
(Meas.Set Menu)" (page 24).
2.
Select "Warm Up" and press the ENT key.
Setting is now possible and the display flashes.
3.
Use the
▲
and
▼
keys to select the warm-up time.
▲
key: Value increases by 1 minute.
▼
key: Value decreases by 1 minute.
The setting range is from 1 minute to 60 minutes.
4.
Press the ENT key to register the setting.
Tip
z
As this instrument uses a UV lamp, a certain amount of time is required for the instructions to
become stable after the power is turned ON.
z
We recommend the warm-up time setting to 30 minutes.
z
When pressing the MENU key, the menu goes back to the previous menu.