5 Troubleshooting
39
5.3.2
Error code, output, and display lamp
When error codes are displayed, output and display lamp will be as follows:
Error code
Control
target
Alarm contact
output
Transmittance
output
Display lamp
Error code display
Maintenance
OFF
Maintenance
ON
System Error
Err
Output error
ON
Hold
Red flash
YES
YES
Hold
Output error
ON
TOC
OFF
Cond
OFF
Temp
OFF
No Sample
Err
Output error
ON
Hold
Red flash
YES
YES
Hold
Output error
ON
TOC
OFF
Cond
ON
Temp
ON
Cond Scale Over
Err
Output error
ON
Hold
Red flash
YES
YES
Hold
Output error
ON
TOC
OFF
Cond
ON
Temp
ON
Temp Scale Over
Err
Output error
ON
Hold
Red flash
YES
YES
Hold
Output error
ON
TOC
OFF
Cond
ON
Temp
ON
Low Pressure
Err
Output error
ON
Hold
Red flash
YES
YES
Hold
Output error
ON
TOC
OFF
Cond
ON
Temp
ON