4 Measurement
30
Q
Option setting for errors (Err. Option Set)
1.
Referring to "
Q
Opening each setting menu" (page 23), display the system setting
screen “Meas. Set Menu”.
With this screen, measurement condition setting item can be selected.
2.
Select “Err. Option Set” and press the ENT key.
The screen display “Err. Option Set” and option setting for errors can be performed.
Selecting the sample water electric conductivity error (Cond Over)
Select between output/not output of abnormal alarm when the conductivity of the sample
water exceeds the upper limit.
Note
Upper limit value for input depends on the setting of “Setting measurement range”.
1.
Perform the procedures 1. and 2. in "
Q
Option setting for errors (Err. Option Set)"
(page 30).
2.
Select “Cond Over” and press the ENT key.
The display flashes to show now the setting can be performed.
3.
Press the
▲
/
▼
key and select for the input electric conductivity error.
▲
key: Yes: Outputs as an error alarm.
▼
key: No: Does not output as an error alarm.
4.
Press the ENT key to confirm the setting.
Reference
For details of “Cond Over”, refer to “ 5 Troubleshooting ” (page 37).
Tip
When pressing the MENU key, the menu goes back to the previous menu.
Setting measurement range
Upper limit
100.0
2.00
µ
S/cm
500.0
10.00
µ
S/cm