3 Installation
21
3.4.5
Attaching terminal covers
Put the terminal covers after completion of wiring of the terminal board.
3.5
Piping configuration
Connect the tube joint to the connection on the bottom of the equipment.
Note
Be sure to hold the connection with a wrench when threading the joint in.
3.6
Adjusting the flow meter
Use the upper needle valve to adjust the flow meter so that its float registers 15 mL/min.
Note
When bubbles adhere within the flow meter''s glass tube or to the float, adjust the flowmeter
after fully opening the needle valve to remove the bubbles.
WARNING
Electric shock
Make sure to put the terminal covers to prevent electric shock.
Sample outlet Rc1/4
(
8 mm o.d./6 mm i.d.:piping connection
)
Sample inlet Rc1/8
(
6 mm o.d./4 mm i.d.:piping connection
)
< Instrument bottom >
Float reading position