4 Measurement
32
4.4.3
Output setting (Output Set)
Q
Alarm contact output setting (Alarm Set)
Set each item for the alarm contact output.
1.
Referring to "
Q
Opening each setting menu" (page 23), display the output setting
screen “Output Set Menu”.
With this screen, the output setting item can be selected.
2.
Select “Alarm Set” and press the ENT key.
The screen displays “Alarm Set Menu” and alarm contact output setting can be per-
formed.
Alarm contact output setting 1 (Alarm1)
1.
Perform the procedures 1. and 2. in "
Q
Alarm contact output setting (Alarm Set)"
(page 32).
2.
Select “Alarm1” and press the ENT key.
The screen displays “Alarm1 Set Menu” and alarm contact output setting 1 can be per-
formed.
Setting item depends on the setting of the marked out part in Table 2.
The part marked with “
○
” is displayed as the setting items.
Table 2 Control target and setting item
Selecting control target (Target)
Select the target to control by alarm contact output.
1.
Perform the procedures 1. and 2. in "
Q
Alarm contact output setting (Alarm Set)"
(page 32).
2.
Confirm “Target” is selected and press the ENT key.
The display flashes to show now the setting can be performed.
3.
Press the
▲
/
▼
key to select the control target.
Err:
Error
Hold: Transmittance output is during Hold
TOC: Total organic carbon concentration
Cond: Electric conductivity
Temp: Temperature
4.
Press the ENT key to confirm the setting.
Note
Control target in “Selecting control target” is to be one item.
Display
Content
Choice
Target
Control target
selection
Err
Hold
TOC
Cond
Temp
Set
Control value
-
-
○
○
○
Delay
Delay time
○
○
○
○
○