Part 5: Controls
Ejecting Films
Use the following procedure if a film seems to be jammed inside the scanner.
X
To remove a jammed film
1
Lift up the interlock (the protective shield on the input tray). If you can see the film,
gently pull it out, using both hands and pulling gently but firmly.
2
If you can’t see the film, or can’t get it out, lift up the scanner control panel cover.
Press the UP and DOWN arrows at the same time and hold them down for 3 seconds.
The film should come out.
3
If the preceding step doesn’t eject the film, power-cycle the system. See ‘
Error!
Reference source not found.
’.
4
If the preceding steps do not fix the problem, call Hologic Technical Support.
Resetting the Clean Scanner Timer
When you touch the
Clean Scanner
button, the system displays:
Clean the scanner as instructed in ‘
. When you finish cleaning
the scanner, touch the
OK
button shown above. When you touch
OK
, the Scanner
Cleaning message on the Maintenance screen changes from ‘Needed’ to ‘OK’.
If you do not clean the scanner, touch the ‘X’ shown above to cancel the timer reset.
Reviewing System Verification Results
The Test Results portion of the Performance screen provides the results of the most
recent system verification testing, including the scanner test.
For more information on the tests, see ‘
68
R2 DM User Manual – PN MAN-01044 Rev 001
Содержание R2 DM
Страница 1: ...User Manual Software Versions 8 6 8 7 PN MAN 01044 Rev 001 R2 DM...
Страница 2: ......
Страница 3: ...Screen Film Mammography R2 DM User Manual Software Versions 8 6 8 7 PN MAN 01044 Rev 001...
Страница 56: ...Part 3 Studies 48 R2 DM User Manual PN MAN 01044 Rev 001...
Страница 68: ...Part 4 Alerts 60 R2 DM User Manual PN MAN 01044 Rev 001...
Страница 92: ...Part 6 Maintenance 84 R2 DM User Manual PN MAN 01044 Rev 001...
Страница 100: ...92 R2 DM User Manual PN MAN 01044 Rev 001...
Страница 105: ......
Страница 106: ......