3.4. Scanning with a Manual Protocol
4
Touch the desired scanning protocol. In the example shown at left,
Screening is selected. The views defined for the selected protocol are
displayed below the protocol list.
Note:
You cannot change the scanning protocol while a case is being
scanned.
5
Touch
Start
. The system status changes from ‘Ready’ to ‘Scanning’, and
the
Start
button changes to
Stop
.
6
A new study icon appears at the top of the study list, and a white circle on
the icon flashes to indicate that the films are being processed. If you touch
this icon, you can monitor the progress of the study as the films are
processed. There are three possible outcomes:
•
Success:
If the system processes the case successfully and verification of
cases is not required, then the system automatically sends the study
results to the output device (display unit, printer and/or PACS
archive). An arrow on the icon flashes to indicate that the system is
sending the results. After the system sends the results, the icon appears
solid green.
•
Verification Required:
If verification of cases is required, then you must
review the scanned image thumbnails and patient information and
confirm that the results are acceptable. If no Alert has been issued, then
the Done button appears on the Studies screen. When you touch Done,
the system sends the results to the output device (display unit, printer
and/or PACS archive).
•
Alerts:
If the study icon is replaced with an Alert icon, you may be
required to take corrective action. In many situations instead of
rescanning the case, you can often correct the alert and reprocess the
images via the Image Correction screen.
For more information, see ‘
3.10. Correcting Scanned Images
’. For
information on alert messages, see ‘
.
7
When the cases have been scanned successfully, the films can be returned
to the patients’ film folders. If necessary, keep the separator sheet with the
case for CAD display.
Note:
If the system has been configured to use a different scanning
protocol by default (on the Controls Scanning screen), then after scanning a
stack of films, the system returns to the default protocol.
R2 DM User Manual – PN MAN-01044 Rev 001
33
Содержание R2 DM
Страница 1: ...User Manual Software Versions 8 6 8 7 PN MAN 01044 Rev 001 R2 DM...
Страница 2: ......
Страница 3: ...Screen Film Mammography R2 DM User Manual Software Versions 8 6 8 7 PN MAN 01044 Rev 001...
Страница 56: ...Part 3 Studies 48 R2 DM User Manual PN MAN 01044 Rev 001...
Страница 68: ...Part 4 Alerts 60 R2 DM User Manual PN MAN 01044 Rev 001...
Страница 92: ...Part 6 Maintenance 84 R2 DM User Manual PN MAN 01044 Rev 001...
Страница 100: ...92 R2 DM User Manual PN MAN 01044 Rev 001...
Страница 105: ......
Страница 106: ......