Part 6: Maintenance
6.3. Cleaning the Scanner
To ensure optimum scanner performance, at least once each day clean the scanner
diffuser, and then clean the scanner’s suction cups using alcohol and the cleaning
swabs provided with the system.
Note:
If you are running less than 80 cases daily, you can clean the scanner once
each day. If you are running 80 or more cases daily, Hologic | R2 recommends that you
clean the scanner twice each day.
To clean the scanner diffuser
1
Place the diffuser cleaner in the input tray, with the cleaning strip facing up, closest to
you.
Diffuser Cleaner
Control Panel
Cleaning Strip
Diffuser Cleaner
2
Lift the cover so you can see the scanner control panel.
3
On the scanner control panel:
a
Press QUIT.
b
Press MENU until ‘4 CLEANING MODE’ appears on the scanner control panel
display, and then press ENTER.
c
Press the UP/DOWN button until ‘DIFFUSER CLEANING’ appears on the
display, and then press ENTER.
d
Press START/STOP. The scanner loads the diffuser cleaner.
e
When you see ‘DIFFUSER CLEANING PULL OUT CLN_FILM’ on the display,
pull out the cleaner sheet, gently but firmly.
4
Repeat steps
1
–
3
two more times.
Note:
If an error occurs, run the diffuser cleaner through the system six times, and
then shut down and restart the system.
5
Store the diffuser cleaner in a safe, clean location.
80
R2 DM User Manual – PN MAN-01044 Rev 001
Содержание R2 DM
Страница 1: ...User Manual Software Versions 8 6 8 7 PN MAN 01044 Rev 001 R2 DM...
Страница 2: ......
Страница 3: ...Screen Film Mammography R2 DM User Manual Software Versions 8 6 8 7 PN MAN 01044 Rev 001...
Страница 56: ...Part 3 Studies 48 R2 DM User Manual PN MAN 01044 Rev 001...
Страница 68: ...Part 4 Alerts 60 R2 DM User Manual PN MAN 01044 Rev 001...
Страница 92: ...Part 6 Maintenance 84 R2 DM User Manual PN MAN 01044 Rev 001...
Страница 100: ...92 R2 DM User Manual PN MAN 01044 Rev 001...
Страница 105: ......
Страница 106: ......