6.3. Cleaning the Scanner
X
To clean the suction cups
1
Place about 5 drops of 70% isopropyl alcohol onto each pad of the suction cup
cleaning plate, and rub it in gently with your finger.
Note:
Use isopropyl alcohol regardless of any instructions on the cleaning plate.
2
Place the suction cup cleaner in the input tray, adjacent to the left side of the tray.
Suction Cup
Cleaner
Fabric Pads
Control Panel
Suction Cup Cleaning Plate
3
Lift the cover so you can see the scanner control panel.
4
On the scanner control panel:
a
Press QUIT.
b
Press MENU until ‘4 CLEANING MODE’ appears on the control panel display,
and then press ENTER.
c
‘SUCTION PAD CLEANING’ appears on the display. Press ENTER to continue.
d
When ‘SET CLN_PLATE & PUSH <START/STOP>’ appears on the display, press
START/STOP.
5
Repeat step
4
two more times.
6
Remove the suction cup cleaning plate from the tray and store it in a safe, clean
location along with the diffuser cleaner.
Caution:
After cleaning the suction cups, wait at least five minutes before scanning
any films. Residual alcohol can damage the films!
7
After cleaning the suction cups, touch the
Controls
tab and then
Performance
.
8
Touch the
Clean Scanner
button. The system displays:
9
Touch
OK
. When you touch OK, the Scanner Cleaning message on the Maintenance
screen changes from ‘Needed’ to ‘OK’.
R2 DM User Manual – PN MAN-01044 Rev 001
81
Содержание R2 DM
Страница 1: ...User Manual Software Versions 8 6 8 7 PN MAN 01044 Rev 001 R2 DM...
Страница 2: ......
Страница 3: ...Screen Film Mammography R2 DM User Manual Software Versions 8 6 8 7 PN MAN 01044 Rev 001...
Страница 56: ...Part 3 Studies 48 R2 DM User Manual PN MAN 01044 Rev 001...
Страница 68: ...Part 4 Alerts 60 R2 DM User Manual PN MAN 01044 Rev 001...
Страница 92: ...Part 6 Maintenance 84 R2 DM User Manual PN MAN 01044 Rev 001...
Страница 100: ...92 R2 DM User Manual PN MAN 01044 Rev 001...
Страница 105: ......
Страница 106: ......