3.10. Correcting Scanned Images
3
Modify the view label, as needed.
To define or change the view label, build it from
left to right: select a side, a view, and then modifiers (as needed).
a
Touch the proper Side selection:
Left
,
Right
, or
Both
(for cleavage view).
b
Touch the proper View selection. Note the two additional selections below the
view area for Partial View and Implant Present.
c
Touch to select Modifiers as needed.
As you build the view label, invalid selections are eliminated. For example, if you
choose Both for the side, the system automatically sets the view to CC and the
modifier to CV. You can chain valid modifiers together as needed.
Some modifiers (for example, M and S) appear before the view designation; other
modifiers (such as ID and TAN) appear after the view designation.
See ‘
’ on page
for a complete list of views and modifiers.
Note:
You must select a view when using the TAN modifier, as TAN can be used with
any view.
Note:
ImageChecker will not process views modified with the Partial View or IP
modifiers. For further information see
Understanding ImageChecker 8.5-8.7.
4
Repeat steps 2 and 3 for each image.
5
Modify the DICOM study description, as needed.
If desired, touch the DICOM Study
Description field and use the keyboard to change the DICOM study description for
the case. The default description is ‘R2 Mammography Digitized Film’.
Note:
All images in the case will contain the same DICOM Study Description;
modifying it in one image changes it for all images in the case.
6
Reprocess the case.
Once the images and their labels are correct, touch
Reprocess
.
The system cancels the old case (removes the old study icon), reprocesses the case (all
images), and creates a new study icon at the top of the study icon list. If the system
does not issue an alert, it:
•
Automatically sends the results to the output devices (PACS, display unit, printer,
etc.) if Verification is off.
•
Waits for you to click the Done button if Verification is on before sending the
results.
R2 DM User Manual – PN MAN-01044 Rev 001
43
Содержание R2 DM
Страница 1: ...User Manual Software Versions 8 6 8 7 PN MAN 01044 Rev 001 R2 DM...
Страница 2: ......
Страница 3: ...Screen Film Mammography R2 DM User Manual Software Versions 8 6 8 7 PN MAN 01044 Rev 001...
Страница 56: ...Part 3 Studies 48 R2 DM User Manual PN MAN 01044 Rev 001...
Страница 68: ...Part 4 Alerts 60 R2 DM User Manual PN MAN 01044 Rev 001...
Страница 92: ...Part 6 Maintenance 84 R2 DM User Manual PN MAN 01044 Rev 001...
Страница 100: ...92 R2 DM User Manual PN MAN 01044 Rev 001...
Страница 105: ......
Страница 106: ......