Hofmann Megaplan Service Manual megaline SSENCE
SSENCE 400/600
Page 49 of 52
6.9 Exchange of personal computer
If the personal computer must be exchanged, please follow these steps:
Save data from old computer (if possible! - If not contact Hofmann Megaplan)
To copy all files use a USB-stick with at least 4 GB memory.
1. copy all calibration files to a USB stick
2. There is a folder named in the same way than the serial number of the machine. this folder contains
factory calibration data.
3. Go to Folder D:\SSENCE\DB and save the H3d.mdb on the stick. Also save personal brand logos.
4. Copy D:\SSENCE\SPECDB.INI to the stick
Go to another computer and download from our homepage:
- Driver for Sentinel Hardware Lock (Dongle)
- PDF-Viewer
- Team Viewer Support module
- Current basic software version
- Current database version
Be sure that the personal computer has:
- 6 USB ports (or more)
- HDMI connection
- Win7 Ultimate installed (use delivered CD to install)
- Two Partitions C: and D:
(SSENCE always is installed in D:)
- Win7 is activated by telephone (manual and help in Windows control panel)
After installation of Windows:
- Install Dongle driver, PDF viewer, Team Viewer Module
- Install SSENCE basic software
- Install SSENCE database
- Copy calibration files to D:\SSENCE
- Copy H3D.mdb to D:\SSENCE\DB
- Copy SPECDB.INI to D:\SSENCE
After first start of SSENCE
- get licence code from Hofmann Megaplan for this machine
- Select currant database in the setup menu.