2.3.4 Possible failure of camera chip:
It seems that we had broken camera modules, in this condition the modules showed incorrect
values in the configuration file CONFIG.INI laying in D:\SSENCE
In the file are values for the cameras:
< This files shows correct values:
This file has abnormal high values. >
If these values are shown camera is black. The camera-chips are not configured well.
Please adjust chips ID left to 0 and ID right to 2. Restart complete machine. (not only
computer) (see 2.3.5) If this is not working try 1 for left and 2 for right and leave the
machine switched off for 5 Minutes. (to decrease possible static electricity.)
- There is a black picture on the screen (F11) Adjusting exposure without result.
- Picture is "foggy", or worse resolution as on the other side-
- Bluring lines on screen (please clean camera lens before to check)
- Different colour (sepia)
- Windows does not recognize the camera although it is mounted correct. (device manager)
The camera housing with
support.
On the back of the camera, the
control chip is mounted (on the
board). On the front side of the
camera
is
a
sticker
which-reflects the programming
of the camera chip, "L" is the
camera is set to the left side of
the camera. ("R" for right) This
default setting can be changed
by a program tool.