5.6 Changing Event Settings
87
Chapte
r 5
Changing
Se
ttings (as
neces
sary)
5
5.6 Changing Event Settings
What is an event?
See:
"Appendix 2 Explanation of Power Supply Quality Parameters and Events"
List of event settings
Event parameter
Order
selection
Additional
functionality
Channel
selection
Threshold (Note 9)
Note
Transient overvoltage
(1,2,3) (4)
(OFF)
0 V peak to 6000 V peak
Specify as absolute value.
1,4
Swell
Slide
(1,2,3) (-)
(OFF)
0% to 200%
1,5,10
Dip
Slide
(1,2,3) (-)
(OFF)
0% to 100%
1,5,10
Interruption
(1,2,3) (-)
(OFF)
0% to 100%
1,5
Inrush current
(1,2,3) (4)
(OFF)
0 A to (varies with range) A
1,4,5
Frequency
200 ms
(U1) (-)
(OFF)
0.1 Hz to about 9.9 Hz
5
Frequency cycle
(U1) (-)
(OFF)
0.1 Hz to about 9.9 Hz
5
Voltage
waveform peak
(1,2,3) (4)
(OFF)
0 V peak to 1200 V peak
1,4,7
RMS voltage
Phase/line
sense
(1,2,3) (4)
(OFF)
0 V to 780 V
Specify upper and lower limits.
1,3,4,5
DC voltage change
(CH4 only)
(-,-,-) (4)
(OFF)
0 V to 1200 V
1,5
Current
waveform peak
(1,2,3) (4)
(OFF)
0 A to (varies with range) A×4
1,4,7
RMS current
Sense
(1,2,3) (4)
(OFF)
0 A to (varies with range) A
1,4,5
DC current change
(CH4 only)
(-,-,-) (4)
(OFF)
0 A to (varies with range) A×4
1,5
Active power
(1,2,3)(sum)
(OFF)
0 to varies with range
Specify as absolute value.
1,4,5,8
Apparent power
(1,2,3)(sum)
(OFF)
0 to varies with range
1,4,5,8
Reactive power
(1,2,3)(sum)
(OFF)
0 to varies with range
Specify as absolute value.
1,4,5,8
Power factor/displace-
ment power factor
PF/DPF
(1,2,3)(sum)
(OFF)
0 to 1
Specify as absolute value.
3,4,5
Negative-phase volt-
age unbalance factor
(-,-,-)
(sum)(OFF)
0% to 100%
5
Zero-phase voltage
unbalance factor
(-,-,-)
(sum)(OFF)
0% to 100%
5
Negative-phase cur-
rent unbalance factor
(-,-,-)
(sum)(OFF)
0% to 100%
5
Zero-phase current
unbalance factor
(-,-,-)
(sum)(OFF)
0% to 100%
5
Harmonic voltage
Orders
0 to 50
Level (RMS)/
content
percentage
(1,2,3) (4)
(OFF)
0 V to 780V/0% to 100%
Specify the 0th order level as an absolute
value.
1,2,3,4,
5,6
Harmonic current
Orders
0 to 50
Level (RMS)/
content
percentage
(1,2,3) (4)
(OFF)
1.3 × (0 to [varies with range]) A /0% to 100%
Specify the 0th order level as an absolute
value.
1,2,3,4,
5,6
Harmonic power
Orders
0 to 50
Level/content
percentage
(1,2,3)(sum)
(OFF)
1.3 × (0 to [varies with range]) W
Specify as absolute value. /0% to 100%
1,2,3,4,
5,6,8
HIOKI PQ3198A961-03
Содержание PQ3198
Страница 2: ...HIOKI PQ3198A961 03...
Страница 6: ...Contents iv HIOKI PQ3198A961 03...
Страница 20: ...Usage Notes 14 HIOKI PQ3198A961 03...
Страница 44: ...2 3 Display Items and Screen Types 38 HIOKI PQ3198A961 03...
Страница 58: ...3 8 Turning the Power On and Off Setting the Default Language 52 HIOKI PQ3198A961 03...
Страница 102: ...5 8 Factory Settings 96 HIOKI PQ3198A961 03...
Страница 116: ...6 4 Displaying Harmonics 110 Holding the display Hold Measured values and waveforms will be held HIOKI PQ3198A961 03...
Страница 198: ...12 4 Downloading Recorded Data to Computer 192 HIOKI PQ3198A961 03...
Страница 264: ...14 4 Disposing of the Instrument 258 HIOKI PQ3198A961 03...
Страница 296: ...A32 HIOKI PQ3198A961 03...
Страница 299: ...Index Index 3 Index Wiring 73 Z Zero adjustment 53 Zero phase A28 HIOKI PQ3198A961 03...
Страница 300: ...Index Index 4 HIOKI PQ3198A961 03...
Страница 301: ...HIOKI PQ3198A961 03...
Страница 302: ...HIOKI PQ3198A961 03...
Страница 303: ...HIOKI PQ3198A961 03...
Страница 304: ...www hioki com HIOKI PQ3198A961 03...